High-voltage
Probes HV-P60/HV-P30
|
TOP
|
|
|
|
|
|
HV-P60
|
|
High-voltage
waveforms of up to DC 60 kV, pulse 80 kV measured by oscilloscope
|
|
|
|
Wide
bandwidth, DC to 50 MHz
|
|
|
|
Maintenance-free,
due to use of silicon rubber - guarantees stable withstand voltage
|
Specifications
|
Maximum
input voltage
|
DC 60
kV or AC peak*
Single-shot pulse of 80 kV, standard pulse wave
|
Bandwidth
|
DC to
50 MHz, within -3 dB
|
Rise
time
|
7 ns
or less (for probes only)
|
Input
RC
|
1000
MW ±5% //5 pF ±2 pF
|
Attenuation
|
2000:1
±5%
|
Cable
length
|
Approx.
4 m
|
Connector
|
BNC
type
|
Used
for
|
Oscilloscope
with input RC: 1 MW, 15 to 50 pF
|
Size
|
Approx.
88 mm (ø) x 338 mm (L)
|
Weight
|
Approx.
3.5 kg (including accessories)
|
Accessories
|
Matching
box, clamp, probe tip, ground lead, adjusting screwdriver, accessory bag
|
Options
|
High-voltage
probe stand SK-301 (produced upon order)
The SK-301 can be used with both the HV-P60 and the HV-P30. It is
attached by turning the probe bracket upside down (for HV-P60/HV-P30).
|
* The
P-P voltage drops if the CW is 100 kHz or greater.
|
|
This is
a passive probe that measures high-voltage waveforms by connecting to an
oscilloscope, and has the following characteristics: the bandwidth is DC to
50 MHz, the maximum input is DC 60 kV, and the attenuation is 2000:1.
|
Characteristic curve
|
|
Maximum Input Voltage and Frequency (Sine Wave) [HV-P60]
|
HV-P30
High-voltage probe stand SK-301 (sold separately)
|
|
High-voltage
waveforms of up to DC 30 kV, pulse 40 kV measured by oscilloscope
|
|
|
|
Wide
bandwidth, DC to 50 MHz
|
|
|
|
Maintenance-free,
due to use of silicon rubber - guarantees stable withstand voltage
|
Specifications
|
Maximum
input voltage
|
DC 30
kV or DC + AC peak*
Single-shot pulse of 40 kV, repetitive signal
|
Bandwidth
|
DC to
50 MHz, within -3dB
|
Rise
time
|
7 ns
or less (for probes only)
|
Input
RC
|
100 MW ±5%
//7 pF or less
|
Attenuation
|
1000:1
±5%
|
Cable
length
|
Approx.
3 m
|
Connector
|
BNC
type
|
Used
for
|
Oscilloscope
with input RC: 1 MW, 15 to 50 pF
|
Size
|
Approx.
66 mm (ø) x 315 mm (L)
|
Weight
|
Approx.
3 kg (including accessories)
|
Accessories
|
Matching
box, clamp, probe tip, ground lead, adjusting screwdriver, accessory bag
|
Options
|
High-voltage
probe stand SK-301 (produced upon order)
The SK-301 can be used with both the HV-P60 and the HV-P30. It is
attached by turning the probe bracket upside down (for HV-P60/HV-P30).
|
* The
P-P voltage drops if the CW is 100 kHz or greater.
|
|
This is
a passive probe that measures high-voltage waveforms by connecting to an
oscilloscope, and has the following characteristics: the bandwidth is DC to
50 MHz, the maximum input is DC 30 kV, and the attenuation is 1000:1.
|
Characteristic curve
|
|
Maximum Input Voltage and Frequency (Sine Wave) [HV-P30]
|
|
|
|